Enable or disable analog output, Serial line output commands, Start measurement output – Vaisala DMT143 User Manual

Page 47: Stop measurement output

Advertising
background image

Chapter 5 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 45

Enable or Disable Analog Output

Use the AOUT command to enable or disable the analog output channel.

AOUT [ON/OFF]<cr>

Example (turn analog output off):

aout off
AOUT : OFF

Serial Line Output Commands

Start Measurement Output

Use the R command to start the continuous outputting of measurement
values as an ASCII text string to the serial line. The format of the
measurement message is set with the FORM command.

R<cr>

Example (measurement message in default format):

r
Tdf= 17.75 'F Tdfatm= -5.25 'F H2O= 954 ppm
Tdf= 17.23 'F Tdfatm= -5.72 'F H2O= 930 ppm
Tdf= 17.06 'F Tdfatm= -5.88 'F H2O= 922 ppm
Tdf= 16.97 'F Tdfatm= -5.96 'F H2O= 918 ppm
...

Outputting the results continues in intervals issued with the command
INTV until stopped.

Stop Measurement Output

You can stop the measurement output by pressing the ESC key.