Figure 26, Wiring of the m12 (8-pin) connector, Table 11 – Vaisala PTB330TS User Manual

Page 67: M12 connector pin assignments

Advertising
background image

Chapter 6 ______________________________ Serial Line Communication With PTB330 Barometer

VAISALA_______________________________________________________________________ 65

To take the user port into use for a serial connection:

-

Connect to the service port first, and configure the user port
communication settings using the SMODE and SERI commands.
The new user port settings will be taken into use when the
transmitter is reset or powered up. For example:

SMODE STOP

SERI 9600 E 7 1

-

Connect to the M12 connector using a serial cable with the proper
wiring. The wiring of the M12 connector is described in

Figure 26

on page 65

. Note that you also have to provide the operation power

for the barometer (10 ... 35 VDC) via the cable.

0503-026

Figure 26

Wiring of the M12 (8-Pin) Connector

Table 11

M12 Connector Pin Assignments

Pin/Terminal

Wire

RS-485 (EIA-485)

1

White

A -

2

Brown

Serial GND

3 (optional)

Green

Serial GND

4

Yellow

-

5

Grey

Supply -

6

Pink

Supply +

7

Blue

B -

8

Shield/Red

Cable shield

Advertising