Vaisala PTB330TS User Manual

Page 84

Advertising
background image

User's Guide ______________________________________________________________________

82 ___________________________________________________________________M210983EN-A

Relative Humidity (HMP155)

Measurement range

0 ... 100 %

Accuracy (incl. non-linearity, hysteresis and repeatability) at

+15 ... +25 °C (+59 ... +77 °F)

±1 %RH (0 ... 90 %RH)
±1.7 %RH (90 ... 100 %RH)

-10 ... +40 °C (+14 ... 104 °F)

±(1.0 + 0.008 x reading) %RH

Factory calibration uncertainty (+20 °C /+68 °F)

±0.6 %RH (0 ... 40 %RH)*
±1.0 %RH (40 ... 97 %RH)*

Humidity sensor

HUMICAP180R
HUMICAP180RC

Response time at +20 °C in still air with a sintered PTFE filter

63 %

20 s

90 %

60 s

* Defined as ±2 standard deviation limits. Small variations possible, see also calibration
certificate.

Temperature (HMP155)

Measurement range

-10 ... +40 °C (+14 ... +104 °F)

Accuracy

-10 ... +20 °C

±(0.176 - 0.0028 x temperature) °C

+20 ... +40 °C

±(0.07 + 0.0025 x temperature) °C

Accuracy over temperature range: see

Figure 28 on page 83

.

Temperature sensor

Pt100 RTD 1/3 Class B IEC 751

Response time with additional temperature probe in 3 m/s air flow

63 %

< 20 s

90 %

< 35 s

Advertising