R starting the continuous output, S stopping the output – Vaisala PTU200 User Manual
Page 30

U
SER
'
S
G
UIDE
_______________________________________________________________________
26 ___________________________________________________________________ M210195
EN
-A
R Starting the continuous output
R <cr>
This command is used to start continuous outputting in the STOP and
RUN modes and after setting the output interval (see page and 18).
Example (in STOP mode):
>reset <cr>
PTU200 / 1.01
>r <cr>
1013.25 hPa 22.9 %RH 22.4 'C
1013.25 hPa 22.9 %RH 22.4 'C
1013.25 hPa 22.) %RH 22.4 'C
...
Example of restarting the outputting (in RUN mode):
>SMODE<cr>
Serial mode
: RUN
>RESET<cr>
1009.1 hPa 42.4 %RH 24.2 'C
1009.1 hPa 42.4 %RH 24.2 'C
...
S<cr>
(
text invisible)
>R<cr>
1009.2 hPa 42.5 %RH 24.2 'C
1009.1 hPa 42.5 %RH 24.2 'C
...
S Stopping the output
S <cr>
This command is used to stop the continuous outputting (activated
with power-up or reset in RUN mode or using the R command).
Example:
>r <cr>
1013.25 hPa 22.9 %RH 22.4 'C
1013.25 hPa 22.9 %RH 22.4 'C
1013.25 hPa 22.9 %RH 22.4 'C
s <cr>
(text invisible)
>