Send mode – Vaisala PTU200 User Manual

Page 34

Advertising
background image

U

SER

'

S

G

UIDE

_______________________________________________________________________

30 ___________________________________________________________________ M210195

EN

-A

SEND mode

In the SEND mode, the transmitter will automatically output one
message at power-up or when triggering the transmitter on using pin 4
of the transmitter. The trigger signal must be a TTL level signal. TTL
HIGH (5 VDC) turns the transmitter ON and TTL LOW (0 VDC)
turns the transmitter OFF.

Examples of outputting with the prompt (ECHO ON) and without the
prompt (ECHO OFF):

>echo <cr>
Echo

:

ON

>smode send <cr>
Serial mode

: SEND

>reset <cr>
1013.25 hPa
>

>echo off <cr>
Echo

:

OFF

reset <cr>

(text invisible)

1013.25 hPa

Example of outputting a minimum amount of (six) characters by
excluding the <cr> and <lf> characters, the pressure unit and the
decimal point with the FORM command:

>form <cr>
4.2 P " " UUUU \r \n
? 6.0 P <cr>
>unit Pa
P unit

: Pa

>smode send <cr>
Serial mode

: SEND

>echo off <cr>
Echo

:

OFF

reset <cr>

(text invisible)

101325
or

(power-up)

101325

Advertising