Chapter 3 ________________________________________________________________ Installation

VAISALA ________________________________________________________________________ 17

CHAPTER 3

INSTALLATION

This chapter provides you with information that is intended to help you

install the MMT310.

Mounting

Selecting a Place for the Probe

Select a place which gives a true picture of the process. Oil should

circulate freely around the sensor; clear oil flow is recommended. Install

the probe directly into the circulation system and not into the oil reservoir

because of deposition.

It is recommended that the sensor head is installed directly in the process

through the ball valve assembly. When the ball valve assembly is used,

the pipe does not have to be emptied or shut down for installation or

removal of the probe. Install the sensor head transversely against the

direction of the process flow.

NOTE

Avoid mounting the transmitter housing close to steam sources or

directly exposed to rain. To ensure an IP 66 class protection, always

mount the transmitter housing with the cable bushings pointing

downwards.

NOTE

Take care not to damage the pipe of the probe. If the pipe is damaged,

the probe head is less tight and will not go through the clasp nut. Make

sure that the filter is tightly fastened to protect the sensors.