Chapter 3 ________________________________________________________________ Installation

VAISALA ________________________________________________________________________ 21

The following two fitting body options are available:

- Fitting Body ISO1/2 solid structure (DRW212076SP)
- Fitting Body NPT1/2 solid structure (NPTFITBODASP)

Table 4

MMT318 Pipe Dimensions

Type

Pipe Dimension

Adjustment Range

Standard

178 mm

120 mm

Optional

400 mm

340 mm

Use teflon tape or thread sealant to seal the connection between the

fitting body and the process/ball valve.

0507-025

Figure 6

Sealing of Fitting Body into Process

Tightening the Clasp Nut

1.

Adjust the probe to a suitable depth according to the type of

installation.

2.

Tighten the clasp nut first manually.

3.

Mark the fitting screw and the clasp nut.

4.

Tighten the nut a further 50 - 60º (ca. 1/6 turn) with a wrench. If

you have suitable torque wrench, tighten the nut to max 45 ± 5 Nm

(33 ± 4 ft-lbs).