SECTION 8. PROCESSING AND PROGRAM CONTROL EXAMPLES

8-23

CR10TCR

ASPTC

(LOWER)

PURPLE

RED

ASPTC

(UPPER)

PURPLE

RED

BLACK

RED

RED

BLACK

RED

BLACK

CLEAR

1H

E3

AG

2H

2L

3H

3L

SWITCHED

12 V CONTROL

C1

G

G

SWITCHED 12 V

JUMPER
WIRE

FIGURE 8.12-1. Connections to Power Two 12 V ASPTCs with the Switched 12 V

on the CR10X Wiring Panel

PROGRAM

*

Table 1 Program

01:

20

Execution Interval (seconds)

01:

If time is (P92)
1:

40--

Minutes (Seconds --) into a

2:

60

Interval (same units as above)

3:

41

Set Port 1 High

02:

If time is (P92)
1:

0

Minutes (Seconds --) into a

2:

1

Interval (same units as above)

3:

30

Then Do

03:

If Flag/Port (P91)
1:

41

Do if Port 1 is High

2:

30

Then Do

04:

Temp (107) (P11)
1:

1

Reps

2:

1

SE Channel

3:

3

Excite all reps w/Exchan 3

4:

1

Loc [ Ref_Temp ]

5:

1

Mult

6:

0

Offset

05:

Thermocouple Temp (DIFF) (P14)
1:

2

Reps

2:

1

±2.5 mV Slow Range

3:

2

DIFF Channel

4:

2

Type E (Chromel-Constantan)

5:

1

Ref Temp Loc [ Ref_Temp ]

6:

2

Loc [ Lower_TC ]

7:

1

Mult

8:

0

Offset

06:

Do (P86)
1:

51

Set Port 1 Low

07:

Z=X-Y (P35)
1:

3

X Loc [ Upper_TC ]

2:

2

Y Loc [ Lower_TC ]

3:

4

Z Loc [ del_TC ]

08:

If time is (P92)
1:

0

Minutes (Seconds --) into a

2:

60

Interval (same units as above)

3:

10

Set Output Flag High

09:

Real Time (P77)
1: 0110

Day,Hour/Minute