CS110 Electric Field Meter

VoltDiff(Internal_RH,1,mV2500,5,True,0,250,0.1,0) 'Internal humidity
measure.

Changing of CS110 desiccant is recommended for internal relative humidity
values

≥80%.

FIGURE 14. Inside of CS110 case illustrating bracket

for holding desiccant

Replacement intervals less than once every six months for the 4
Unit (PN: 4905) desiccant pack within the sealed CS110 case
indicate a problem with the CS110 case seal or with the
desiccant packs being used.

NOTE

Procedure for Changing Desiccant:

1. Remove the CS110 case lid by unscrewing the captive screws that attach

the lid to the main body of the CS110.

38