Figure g-4. determination of csite g-5 – Campbell Scientific CS110 Electric Field Meter User Manual
Page 77
Advertising
![background image](/manuals/415899/77/background.png)
Appendix G. CS110 2 Meter CM10 Tripod Site
Site Correction for Tripod CS110 Only Site - October 2, 2005
1 minute average of 1 second data
Results indicate Csite = 0.105.
y = 0.1051x - 35.664
R
2
= 0.9996
-10000
-8000
-6000
-4000
-2000
0
2000
4000
6000
8000
-80000
-60000
-40000
-20000
0
20000
40000
60000
80000
Uncorrected (Csite = 1) Electric Field (volt/meter) for Tripod CS110 Only site
SN: 1023 Mparallel_Plate = 81.77 volt/meter*millivolt
Elec
tri
c Fie
ld (vol
t/
m
ete
r
fo
r U
p
ward Faci
ng
C
S
110
S
N
:1022 Mparallel_plate
=
87.6 volt
/met
er*mi
llivolt
Tripod CS110 Only Uncorrected E_Field 1_Min_Avg volts/m
Linear (Tripod CS110 Only Uncorrected E_Field 1_Min_Avg
volts/m)
FIGURE G-4. Determination of Csite
G-5