Appendix D. Servicing the CS110
The CS110 has been designed to provide reliable electric field measurements
and to simplify and minimize maintenance. An exploded view illustrating the
various major components of the CS110 is illustrated in Figure D-1.
Desiccant Holder Bracket
Case Lid
FIGURE D-1. Exploded view of CS110 Electric Field Meter
D.1 Lid Gasket
Whenever removing the CS110 case lid inspect the lid gasket for damage and
replace (PN: 17533) in order to form a good seal.
D.2 Changing Out the CR1000
The CR1000 datalogger is accessible in the CS110 by removing the case lid.
The CR1000 is secured to the CS110 via the desiccant holder bracket along with
the two thumb screws on the CR1000 module. Light prying on the CR1000
module is required to unplug the three 40-pin connectors that interface to the
Motor Assembly
CR1000 Datalogger
CS110 Panel Board
Interface Plate
Lid gasket
CS110 Case