Campbell Scientific CSIM11 pH and ORP Sensors User Manual

Page 11

Advertising
background image

CSIM11 pH and ORP Probes

5. Under the

Available Sensors and Devices list, select the Sensors |

Temperature folder. Select 107 Temperature Probe. Click

to

move the selection to the

Selected device window. Use the data default of

degree Celsius.

6. Under the

Available Sensors and Devices list, select the Sensors | Water |

Quality folder. Select CSIM11 pH Probe. Click

to move the

selection to the

Selected device window. Click on the Solution

temperature (Deg C) reference box and select T107_C. The probe

usually requires an offset adjustment. The white panel at the bottom of the

Properties window provides a procedure for determining the value that

should be entered in the

Offset box.