C.2 tips and techniques – Campbell Scientific CSIM11 pH and ORP Sensors User Manual

Page 32

Advertising
background image

Appendix C. Detailed Calibration Procedure and Manufacturer Tips

7. Remove and rinse the electrode, then place it in the 10.00 buffer.

8. In the datalogger real time monitor mode, the electrode should read

10.00 pH, ±0.2, depending on temperature.

9. If the readings in Steps 6 and 8 show a slope error (error proportional to

the difference between the buffer pH and 7), then the probe has a slight

slope error, which should be easy to correct by adjusting the multiplier in

VoltDiff() or Volt (Diff) (P2).

C.2 Tips and Techniques

Stirring the buffers and samples improves the stability and speed of response of

the measurement.

Rinse the electrode with distilled water between samples and lightly blot the

water on a paper towel before immersing it in the next sample. Never wipe the

pH bulb since dust may scratch the delicate gel layer impairing response.

Rinsing the electrode with a small amount of the sample before immersing it

will eliminate any contamination of the sample.

The simplest form of temperature compensation compensates for the change in

the electrode’s slope due to temperature, in accordance with the Nernst factor.

It does not compensate for changes in the actual pH of the sample that occurs

with a change in temperature. The pH of a sample at 25 °C is most likely

different than the pH of that same sample at 75 °C.

The temperature of the electrode, the sample, and the calibration buffers should

be the same. Allow the electrode to come to temperature equilibrium with the

sample before recording the measurement value. Measurements made more

than 20 °C from the calibration temperature should include a one-point

calibration at that temperature for maximum accuracy.

Low ionic strength samples, highly viscous samples, and salt brines slow the

speed of response of the electrode. While the electrode will be stable to a

change in pH buffers after 10 to 15 seconds, it may take up to five minutes for

the electrode to stabilize in a difficult sample.

C-2

Advertising