CSIM11 pH and ORP Probes

sensor in 3.5 molar KCl, or water if KCl is not available, and heat to

approximately 60 °C for 15 minutes; retest.

SYMPTOM: Slow response and/or noisy, erratic readings.

POSSIBLE REASON: Slow response is caused by a very high impedance in

either glass or reference electrode.

CHECK:

1. Visually inspect pH bulb and reference junction for coating or clogging. If

coated, clean as described in Section 9.2, Electrode Cleaning.

2. Allowing a pH electrode to dry out raises the impedance dramatically.

Soaking the electrode in 0.1 normal HCl for 1/2 hour and rinsing with

distilled water should speed response (always keep electrodes wet).

3. Chemical degradation of pH glass can occur rapidly in a high temperature

or high pH environment. This also increases impedance yielding sluggish

response.

4. Low temperature environments also increase impedance, with impedance

doubling every 8 °C temperature drop from 25 °C.

5. Test wiring for intermittent connections by moving wires at the connector

on the datalogger wiring panel. Tighten as necessary. This is a common

source of erratic signals.

SYMPTOM: No response to pH change.

POSSIBLE REASON: A cracked glass bulb.

CHECK:

If electrode reads between 5.8 and 6.2 pH in all solutions, visually inspect glass

bulb for cracks. Discard probe if it is cracked.

POSSIBLE REASON: Short circuit.

CHECK:

If electrode reads 7.0 pH or 0.0 mV when connected to the datalogger, visually

inspect cable for damage. If there is no visible damage, test electrode in 4.0

buffer.