C.1 calibration, Detailed calibration procedure and – Campbell Scientific CSIM11 pH and ORP Sensors User Manual

Page 31

Advertising
background image

Appendix C. Detailed Calibration

Procedure and Manufacturer Tips

C.1 Calibration

Good laboratory practices should be used and protective

gloves and safety glasses should be worn while handling

any solvents or chemicals. If you are unsure of the proper

technique for handling a chemical or of its hazardous

properties, it is best to discard the electrode eliminating the

risk of danger.

MATERIALS

Buffers
a) 4.01 pH (potassium bipthalate)

b) 7.00 pH (potassium phosphate)

c) 10.00 pH (sodium borate and carbonate)
Thermometer

Beakers

Datalogger programmed to read the pH probe

PROCEDURE

1. Fill beakers with appropriate buffers and continuously measure the

temperature of the buffer in use. If possible, temperature should be at

equilibrium (probe and buffers) before continuing with calibration.

2. Calculate the Nernst temperature compensation for the probe’s current

temperature and adjust the multiplier in the differential voltage instruction

(

VoltDiff() in CRBasic or Volt (Diff) (P2) in Edlog) to the appropriate

value. The slope change is usually taken to be –0.2 pH/mV/°C. Examples

of adjusted multipliers would be the multiplicative inverse of the following

slopes: –58 mV/pH at 20 °C, –59 mV/pH at 25 °C, and -60 mV/pH at 30 °C.

3. Zero the pH probe (pH 7 corresponds to 0.0 mV at 25 °C) by placing the

electrode in a 7.0 buffer with the probe connected to the datalogger. Place

the datalogger in real time monitor mode. Electrode should read 7.0 (or

whatever the solution pH should be at the present temperature) ±0.2 pH.

4. Using the offset in the differential voltage instruction (

VoltDiff() or Volt

(Diff) (P2)), adjust the datalogger to read 7.0 pH, or whatever the solution

pH should be at the present temperature.

5. Remove the probe from the pH buffer, rinse the electrode with distilled

water, and place in 4.01 buffer.

6. Place datalogger in real time monitoring mode. The electrode should read

4.01 ±0.2, depending on temperature.

CAUTION

C-1

Advertising