3 wiring, 4 datalogger programming, Wiring – Campbell Scientific CSIM11 pH and ORP Sensors User Manual

Page 18: Datalogger programming, 1. csim11/csim11-orp wiring, 2. 107 wiring

Advertising
background image

CSIM11 pH and ORP Probes

7.3 Wiring

The CSIM11 and CSIM11-ORP are connected to differential channels.

Connections to Campbell Scientific dataloggers are given in TABLE 7-1.

TABLE 7-1. CSIM11/CSIM11-ORP Wiring

Color

Description

CR9000(X)

CR5000

CR3000

CR1000

CR800

CR850

CR510

CR500

CR10(X)

21X

CR7

CR23X

Red

Signal

Differential

High

Differential

High

Differential

High

Green Signal

Reference

Differential

Low

Differential

Low

Differential

Low

Brown Signal Ground

AG

Often Campbell Scientific’s 107 temperature probe is used to compensate for

thermal effects. TABLE 7-2 shows the connections to the 107 probe.

TABLE 7-2. 107 Wiring

Color

Description

CR9000(X)

CR5000

CR3000

CR1000

CR800

CR850

CR510

CR500

CR10(X)

21X

CR7

CR23X

Black Voltage

Excitation

EX or VX

EX

EX

Red

Temperature

Signal

Single-ended

channel

Single-ended

channel

Single-ended

channel

Purple Signal Ground

AG

Clear

Shield

G

7.4 Datalogger Programming

Short Cut is the best source for up-to-date datalogger programming code.

Programming code is needed,

when creating a program for a new datalogger installation

when adding sensors to an existing datalogger program

If your data acquisition requirements are simple, you can probably create and

maintain a datalogger program exclusively with Short Cut. If your data

acquisition needs are more complex, the files that Short Cut creates are a great

source for programming code to start a new program or add to an existing

custom program.

10