Appendix E. Menu System Map

WELCOME TO THE PWS100 SETUP MENU

ID {id}
SN {sn}

(0) message 0
(1) message 1
(2) message 2
(3) set time & date
(4) configuration
(5) password
(6) weather & alarm parameters
(7) terminal (any changes will be lost)
(8) info
(9) done

Make a selection: {input}

CONFIGURATION

(1) PWS100 ID
(2) TRH probe type
(3) Wetness probe type
(4) Aux probe type
(5) Hood heater temperature (C)
(6) Dew heater mode
(7) Output mode
(8) Calibration

Make selection: {input}

4

Menu operation
1..Menu will activate when terminal is open
2..Return only will go to previous menu
3..Incorrect entry will return to same menu
4..Delete will clear current entry
5..Esc will clear current entry and go back to previous menu

PWS100 Calibration

CAUTION! INCORRECT ENTRIES CAN SERIOUSLY DEGRADE ACCURACY

to edit or view instrument calibration, type 'confirm' : {input}

PWS100 View / Enter calibration disc constants

s/n 1234
velocity (m/sec) 5.000
size (mm) 2.386
visibility (mV) 0.318

enter s/n 1234 : {input}
enter size constant (mm) 2.386 : {input}
enter visibility constant (mV) 0.318 : {input}

0

PWS100 View / Adjust calibration

Latest results from calibrator date 21:05:08

latest target--range calibration date

Velocity_Centre 5.000 4.900 5.100 21:05:08
Velocity_Left 5.000 n/a n/a n/a
Velocity_Right 5.000 n/a n/a n/a
Size_Centre 2.350 2.311 2.461 21:05:08
Visibility (mV) 0.300 0.286 0.350 21:05:08
Visibility Offset (uV) 0.000 -2.000 2.000 n/a

(0) no change
(1) adjust calibration (THIS WILL TAKE IMMEDIATE EFFECT)

Make selection : {input}

Update velocity calibration y/n : {input} saved
Update size calibration y/n : {input} saved
Update visibility calibration y/n : {input} saved

1

8