Using ground mode, Using ground mode -12 – Agilent Technologies B2201A User Manual

Page 110

Advertising
background image

4-12

Agilent B2200 User’s Guide, Edition 2

Programming
Programming Examples

Using Ground Mode

The following example uses the ground mode.

Figure 4-4

Ground Mode Example

Setup:

•

Channel configuration mode: Auto

•

Connection rule: Single. Multiple connection is available for the ground port.

•

Connection sequence: Break_Before_Make

•

Used module: All switch modules installed in the mainframe.

•

Ground port: AUX Input 12

•

Ground enabled input port: SMU Input 5 to 8

•

Ground enabled output port: All output ports

•

Connection paths (connected when the ground mode is ON):

from ground port to all ground enabled input/output ports

•

Connection paths (connected by the :ROUT:CLOS command):

The following output ports will be disconnected from the ground port, and the
following input-output connections will be made.

from SMU1 to Output 1 (channel list 00101)

from SMU2 to Output 2 (channel list 00202)

from SMU3 to Output 3 (channel list 00303)

from SMU4 to Output 4 (channel list 00404)