Agilent Technologies B2201A User Manual

Page 180

Advertising
background image

5-54

Agilent B2200 User’s Guide, Edition 2

SCPI Command Reference
Status Reporting Structure

Figure 5-1

Status Reporting Structure of B2200

The above figure shows Service Request Generation. When a condition occurs that
requires service, the instrument sets Request Service bit (RQS

− bit6) of Status

Byte, and sends a Service Request (SRQ) via GPIB bus to the controller. So, the
controller can execute an interrupt service routine (must be in the program) that uses
a Serial Poll to read Status Byte of each instrument to determine which instrument
requested service. See “Status Byte Register” on page 5-55.

Advertising
This manual is related to the following products: