Agilent B2200 User’s Guide, Edition 2

4-15

Programming

Programming Examples

Table 4-6

Couple Mode Example

Imports Agilent.TMFramework
Imports Agilent.TMFramework.DataAnalysis
Imports Agilent.TMFramework.DataVisualization
Imports Agilent.TMFramework.InstrumentIO
Module Module1

Sub Main()

Dim B2200 As New DirectIO("GPIB0::22::INSTR")
Dim channels As String = "(@101,202,303,505)"

'8

B2200.WriteLine("*RST")
B2200.WriteLine(":ROUT:FUNC ACON")

'10

B2200.WriteLine(":ROUT:CONN:RULE ALL,SROU")
B2200.WriteLine(":ROUT:CONN:SEQ ALL,BBM")
B2200.WriteLine(":ROUT:COUP:PORT ALL,'3,5'")

'13

B2200.WriteLine(":ROUT:COUP:STAT ALL,ON")
B2200.WriteLine(":ROUT:CLOS " & channels)

'15

MsgBox("Click OK to start measurement.", vbOKOnly, "")
Console.WriteLine("Measurement in progress. . ." & Chr(10))

'insert the code for measurement

B2200.WriteLine(":ROUT:COUP:STAT ALL,OFF")

'19

B2200.WriteLine(":ROUT:OPEN:CARD ALL")
B2200.Close()
MsgBox("Click OK to stop the program.", vbOKOnly, "")
Console.WriteLine("Measurement completed." & Chr(10))

End Sub

'24

End Module

Line

Description

8

Defines the channel list in the channels variable (string).

9 to 12

Resets the Agilent B2200. And sets the Auto configuration mode, the Single connection
rule, and the Break_Before_Make connection sequence.

13 to 14

Sets the couple ports, and turns the couple mode ON.

15

Closes switches to connect the input-output paths specified by channels.

15 to 25

Almost same as the code shown in the lines 13 to 22 of Table 4-3. The command used to
turn the couple mode OFF has been inserted to the line 19.