5-18

Agilent B2200 User’s Guide, Edition 2

SCPI Command Reference
Command Summary

DIAG subsystem

:DIAGnostic subsystem has commands for executing the self-test function. For
more info, see “Selftest Menu” on page 3-34.

The :DIAGnostic subsystem commands ignore the B2200 channel configuration
mode. For :DIAG commands that require a card number, you specify 1, 2, 3, 4, or
ALL.

Command

Description

:DIAG:TEST:CARD:CLE card_number

Clears relay test result (pass/fail) of the specified card.

card_number: 1, 2, 3, 4, or ALL

:DIAG:TEST:CARD[:EXEC]? card_number

Executes relay test, then returns result: 1 (fail card exists), 0
(pass).

card_number: 1, 2, 3, 4, or ALL

:DIAG:TEST:CARD:STAT? card_number

Returns most recent relay test result: 1 (fail), 0 (pass), -1 (not
tested).

card_number: 1, 2, 3, 4

:DIAG:TEST:FRAM:CLE item

Clears specified test result.

item: CONT (controller test), FPAN (front panel interface test),
LED, PEN, or BEEP

:DIAG:TEST:FRAM[:EXEC]? item

Executes specified test, then returns test result: 1 (fail), 0
(pass).

item: CONT (controller test), FPAN (front panel interface test),
LED, PEN, or BEEP

:DIAG:TEST:FRAM:STAT? item

Returns most recent test result of the specified test: 1 (fail), 0
(pass), -1 (not tested).
item: CONT (controller test), FPAN (front panel interface test),
LED, PEN, or BEEP