7-14

Agilent B2200 User’s Guide, Edition 2

Error Messages
B2200 Specific Error Messages

3026

Cannot use same port for Unused and Couple

The couple mode and the ground mode cannot be used in parallel
when a couple port and the ground enabled input port (unused port)
have been assigned to the same input port.

3027

Cannot use Unused Port during Auto Ground Mode ON

The ground enabled input port (unused port) must be opened during
the ground mode is ON.

3030

Bad input port number

The input port number must be 1 to 14.

3031

Bad output port number

The output port number must be 1 to 12, 24, 36, or 48. Maximum
number depends on the switch module configuration.

3032

Bad setting memory number

The internal memory number for the setup data must be 1 to 8.

3033

EEPROM programming failure

Cannot change the GPIB address or update internal memory data.
Contact your nearest Agilent Technologies service center.

3034

EEPROM reading failure

Cannot read the GPIB address or internal memory data. Contact your
nearest Agilent Technologies service center.

3035

Setting memory data is invalid

Cannot read the setup data in the internal memory. The data has been
broken. Delete the data.

3036

Cannot load this setting data in this configuration

Cannot read the setup data in the internal memory due to the
configuration mismatch. Delete the data or leave it until restoring the
module configuration.

Error

Number

Error Message and Description