For troubleshooting – Agilent Technologies N9360A User Manual
Page 32
Advertising

5-4
N9360A Multi UE Tester Quick Reference Guide
5
W-CDMA System
For Troubleshooting
Figure 5-3 [Manual Test: Stand-by] Screen
1
Press the Manual Test softkey on the [Initial] screen.
2
Move the cursor to the “Procedure” field and select a
pre-defined test procedure.
3
Turn the mobile phone on.
4
Wait for the completion of Location Update.
5
Select “RMC” for BS Call at the “BS Call” input field.
6
Press the BS Call softkey.
7
While “Connection (RMC)” is highlighted in the test flow,
press the Trigger softkey to start single measurement or the
Trigger Sing/Cont
softkey to start continuous measurement.
Press the Trigger Sing/Cont softkey again to terminate
continuous measurement.
8
Press the Release softkey to finish the Test.