NOVUS TXRAIL0 - 10VDC User Manual

NOVUS Sensors

Advertising
background image

INSTALLATION

The transmitter is intended for DIN rail mounting. Its drawing is presented in Figure 04.

Figure 04 – Transmitter dimensions

ELECTRICAL CONNECTIONS

Figure 05 below shows the transmitter connections to the sensor and power supply.

Terminals 1, 2 and 3 are used for sensor input. For 2-wire Pt100, terminals 2 and 3
shall be connected together.

Figure 05 – Transmitter connections – Pt100

Figure 06 – Transmitter connections – Thermocouple

It is important to follow the recommendations below:

Signal wires should be installed in grounded conduits and away from power or

contactor wires.

The instrument should have its own power supply wires, which should not be

shared with electrical motors, coils, contactors, etc.

Installing RC filters is strongly recommended at contactor coils or any other

inductors.

System failure should always be taken into account when designing a control

panel to avoid irreversible damage to equipment or people.

SPECIFICATIONS

Sensor input: User defined. The supported sensors are lited in Table 01, along

with their maximum ranges.

Thermocouples: Types J, K, R, S, T, N and E, according IEC 60584-1 (ITS-90.

Impedance >> 1 M

Pt100:

Excitation: 0.18 mA, 2 or 3-wire connection (for 2-wire

sensors, tie terminals 2 and 3 together).

= 0.00385, according to IEC 60751 (ITS-90)

Voltage: 0 to 50 mVdc. Impedance >> 1 M

Sensor Type

Range

Minimum measurement span

Thermocouple K

0 to 1370 °C

100 °C

Thermocouple J

0 to 760 °C

100 °C

Thermocouple R

0 to 1760 °C

400 °C

Thermocouple S

0 to 1760 °C

400 °C

Thermocouple T

0 to 400 °C

100 °C

Thermocouple N

0 to 1300 °C

100 °C

Thermocouple E

0 to 720 °C

100 °C

Pt100

-200 to 650 °C

40 °C

Voltage

0 to 50 mV

5 mV

Table 01 – Transmitter input sensors

Output: 0 to 10 Vdc, linear with respect to the measured temperature.
Total Accuracy: better than 0.3 % of the maximum range for thermocouples and

0.2 % for Pt100 and voltage (for voltage output above 100 mV).
Resolution: 0.025 V (12 bits)
Maximum Current (Load): 2 mA
Response Time: < 100 ms
Power supply: 18 to 35 Vdc
Operation Temperature: -40 to 85 °C
Humidity: 0 to 90 % RH
Protection: IP40
Electromagnetic compatibility: EN 50081-2, EN 50082-2
No isolation between input, voltage output and power supply.
Internal protection against polarity inversion.
Cold junction compensation for thermocouples.
Housing: ABS plastic. Dimensions: Refer to Figure 04.

T E M P E R A T U R E T R A N S M I T T E R

T X R A I L 0 - 1 0 V D C

O P E R A T I N G M A N U A L


































Man 5001496

WARRANTY

Warranty conditions are available on our web site

www.novusautomation.com

.

Advertising