30

8.

Press

. “al alternating with “tds” briefly displays along

with the “PPm” value of the sample. “P

r

es cal” will display

when the reading stabilizes.

9.

Press cal to accept and advance to the first titration prompt.

add” alternating with “a1” displays; then “agit t1” alternating

with “hOld” displays. (The number after “t” indicates which

titration point you are on. The number “1” here indicates this is

the first titration point. Subsequent titration points are sequenced

numerically: 2, 3, 4, etc.)

10.

Using the pipette, add 100 µL of Reagent: A1, close the cell

extender cap securely, then agitate and hold. Repeat as

prompted. “P

r

es cal” will display.

11.

Press cal to advance to the next titration point. “add

alternating with “a1” displays; then “agit t2” alternating with

hOld” displays.

12.

Tap the cell extender cap to dispel any solution clinging to the

cap back into the cell. Carefully open the cell extender cap so

as not to spill any solution.

13.

Continue adding 100µL of Reagent A1, agitating and holding the

unit as prompted, then pressing cal as prompted. An alkalinity

value will display when the titration measurement is complete.

14.

Press cal. The “cal” icon will display.

15.

Adjust the displayed value to “100” using the UP or DOWN key.

The unit will not allow you to scroll up or down to “100” if it

is beyond ±10% of the reading. Ref. Total Return to Factory

Settings “FAC SEL”, pg. 41 to reset the Alkalinity Calibration to

the “FAC” setting along with all other calibrations.

16.

Press cal to accept. The unit will automatically power off after

the period of inactivity defined in the AUTO OFF setting (ref.

AUTO OFF, pg. 43).

6. Hardness Calibration

Prepare the materials required for titration: Hardness Standard Solution

200PPM, Reagent: C1, Reagent: H1, Reagent: H2, Reagent: H3, cell

extender, T-plunger, 100 µL pipette with a clean tip installed, and two

extra clean tips.
nOte: If you do not complete each titration point within the 3-minute

timeout period, the unit exits the Hardness function. Press harD to

start the titration process over from step 1. Between titrations, the unit

will automatically power off after the period of inactivity defined in the

AUTO OFF setting (ref. AUTO OFF, pg. 43).