Myron L 9PTK User Manual

Page 46

Advertising
background image

42

2.

Press

to display the stored memory records.

3.

Press

repeatedly until you pass the “cl

r

all” and the

C-F locations. The display will show a “Fac sel” (see Figure 20).

4.

Press

to accept the resetting. Display will return to

Conductivity.

Xiii.

cell checK

The cell check verifies the cleanliness of the conductivity/TDS/

resistivity sensor. In normal use the cell may become dirty or coated

and require cleaning. If the display is showing “.00” when the cell cup

is dry, the sensor is probably clean. However, when testing high purity

water in resistivity (“res”) mode improved accuracy may be desired.

No matter what a manufacturer claims, a sensor can and will become

contaminated or coated and, therefore, require cleaning. A true 4-wire

sensor, as in the Ultrameter III, helps to mitigate contamination, but

NO SENSOR IS 100% IMMUNE.

1.

Press

.

2.

Press

to display the stored memory records.

3.

Press

repeatedly until you pass the FAC SEL location.

The display will show a “cell ch” (see Figure 21).

4.

Press

to test. If cell is clean, “good” will momentarily be