Myron L 9PTK User Manual

Page 50

Advertising
background image

46

If none of the Solution Selection icons are displayed (i.e., KCl,

NaCl or 442), nothing has been linked to User mode.

6.

Press

key to accept the setting. Pressing any of the

measurement keys will exit without changing the setting. User

mode “linc” is now complete. The User mode will now use

the calibration gain constant used for the calibration of the

Standard Solution as outlined above.

C. Canceling User mode Calibration “Linc”

The Ultrameter III must be in User linked mode in order to cancel the

linc” (ref. Solution Selection, pg. 22).

1.

Press “Linked” measurement key

,

or

.

Two solution icons will be shown in the left side of display —

User” and another, e.g., “Kcl”.

2.

Press

key until the menu “linc” appears (see Figure 27).

3.

Press

key; the instrument will display both “sel” and

the “User” Icon.

4.

Press

key until “User” is the only solution icon being

displayed.

5.

Press

key.

6.

The User mode calibration “linc” has now been canceled.

nOtes:

1.

To maintain repeatability, use the same standard solutions for

future calibrations.