Company calls “natural water”. These are modeled in a mixture called

“442™” which the Myron L Company markets for use as a calibration

standard, as it does standard KCl and NaCl solutions.

The Ultrameter III contains algorithms for these 3 most commonly

referenced compounds. The solution type in use is displayed on the

left. Besides KCl, NaCl, and 442, there is the “User” choice. The

benefit of “User” is that one may enter the temperature compensation

and TDS ratio by hand, greatly increasing accuracy of readings for a

specific solution. That value remains a constant for all measurements

and should be reset for different dilutions or temperatures.

C. When does it make a lot of difference?

First, the accuracy of temperature compensation to 25°C determines the

accuracy of any TDS conversion. Assume we have industrial process

water to be pretreated by RO. Assume it is 45°C and reads 1500 µS

uncompensated.

1.

If NaCl compensation is used, an instrument would report 1035

µS compensated, which corresponds to 510 ppm NaCl.

2.

If 442 compensation is used, an instrument would report 1024

µS compensated, which corresponds to 713 ppm 442.

The difference in values is 40%.

In spite of such large error, some users will continue to take data in

the NaCl mode because their previous data gathering and process

monitoring was done with an older NaCl referenced device.

Selecting the correct Solution Type on the Ultrameter III will allow the

user to attain true TDS readings that correspond to evaporated weight.

If none of the 3 standard solutions apply, the User mode must be used.

Temperature Compensation (Tempco) and TDS Derivation below, details

the User mode.

XXii. temPeratUre cOmPensatiOn (tempco) and tDs

DerivatiOn

The Ultrameter III contains internal algorithms for characteristics of the

3 most commonly referenced compounds. The solution type in use is

displayed on the left. Besides KCl, NaCl, and 442, there is the User

choice. The benefit of User mode is that one may enter the tempco and

TDS conversion values of a unique solution via the keypad.