Faults / 11. dismounting, return and disposal, Faults, Dismounting, return and disposal – WIKA 70-8xx User Manual

Page 14

Advertising
background image

GB

14

11472988.02 04/2013 GB/D/F/E

WIKA operating instructions expansion thermometer, model 70

10. Faults

Faults

Causes

Measures

The pointer drops below the

start of the scale range.

Damaged capillary

Carefully dismount the instrument,

pack it (tightly sealed) and return it

to the manufacturer for repair.

Instrument switches at the

wrong temperature

Switch point is set incorrectly.

Check the switch point

Instrument does not switch

Defect in switch setting/ micro

switch

Return the instrument to the

manufacturer for repair

CAUTION!

If faults cannot be eliminated by means of the measures listed above, take the

instrument out of service immediately, ensure that no pressure/temperature

or signal are still present and secure the instrument from being put back into

operation inadvertently.

In this case, contact the manufacturer.

If a return is needed, please follow the instructions given in chapter

11.2 "Return".

11. Dismounting, return and disposal

WARNING!

Residual media at the dismounted instrument can result in a risk to persons,

the environment and the equipment.

Take sufficient precautionary measures.

11.1 Dismounting

WARNING!

Risk of burns!

Let the instrument cool down sufficiently before dismounting!

During dismounting there is a risk of dangerously hot pressure media

escaping.

Only disconnect the expansion thermometer once the system has been depressurised!