Control mode, Pneumatic high-speed pressure controller cpc3000 – WIKA CPC3000 User Manual
Page 28

Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1
28
11498171.01 08/2009 GB
[CONTROL]:
In CONTROL mode the instrument provides a precise pressure output (equal to the setpoint
+/- the stability specification) at the
MEASURE/CONTROL port. The indication of the
current pressure value will turn green when the setpoint has been reached and the stable
window settings have been satisfied. "Figure - CONTROL Mode" shows the state of the
isolation valves in measure mode. Notice that the regulator is active in the CONTROL
mode.
Figure - CONTROL Mode
CONTROL Mode
Valve is open
Valve is closed
SUPPLY
VENT
Optional barometric
reference
Optiona
barometric
sensor
Pressure
transducer
REFERENCE
MEASURE / CONTROL
EXHAUST / VACUUM
Regulator
is active