Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1
48
11498171.01 08/2009 GB
Command
Data
Response/Function
Vent
None
Instrument placed in Vent Mode.
Vent?
<sp>(Yes or No)<cr><lf>
Returns Yes if instrument is in Vent, No
if otherwise.
Window
Value in current units
Sets the exponential filter window for
the active sensor.
Window?
<sp>n.nnnnne+nn<cr><lf>
Returns the exponential filter window
for the active sensor.
Zero
desired pressure or ?
Sets zero to set pressure or for ?,
clears previous value.
Zero?
<sp>xxxxxxx<cr><lf>
Returns zero offset for active trans-
ducer.
8.5.2 PCS 400 Commands Emulated
Command
Data
Response/Function
_pcs4 autorange <value>
0 or 1
1 turns autorange on, 0 – off.
_pcs4 autorange?
Returns 1 if in autorange, 0 if in
range hold.
_pcs4 cal a/d
Not applicable.
_pcs4 cal atm
Performs pcs400 1 pt cal.
_pcs4 cal span <value>
Sets the span of the active trans-
ducer to <value>.
_pcs4 cal zero <value>
Sets the zero of the active trans-
ducer to <value>.
_pcs4 cal_disable_off
Enables zero or span calibrations if
previously disabled.
_pcs4 cal_disable_on
Prevents zero or span calibrations.
_pcs4 ctrl <value><sp><unitno>
Sets control value – will take effect
immediately if instrument is in
control mode.
_pcs4 ctrl?
Returns the current control point in
current engineering units.
_pcs4 ctrlmax <value>
Sets maximum control value.
_pcs4 ctrlmax?
Returns current maximum control
pressure.
_pcs4 ctrlmin <value>
Sets minimum control value.
_pcs4 ctrlmin?
Returns current minimum control
pressure.
_pcs4 emul?
Returns ptype emulation mode.
_pcs4 default
Sets default values into instrument.
_pcs4 err?
Returns the error number and
description.
_pcs4 exhaustp?
Returns exhaust pressure.