Pneumatic High-Speed Pressure Controller CPC3000

WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1

45

11498171.01 08/2009 GB

Command

Data

Response/Function

IP?

<sp>nnn.nnn.nnn.nnn<cr><lf>

Returns the IP address of the instru-
ment.

Keylock

Yes or No

Locks or unlocks keyboard.

Keylock?

<sp>(Yes or No)<cr><lf>

Returns Yes or No.

List?

<sp>Pri,X,X;Sec,X,X;Bar,1<cr><lf>

Returns list of available turn-downs on
installed sensors in the active channel.
X will be non-existent if the turndown
isn’t available.

Listrange?

PRI,1, min,max,2, min, max;SEC,1,
min,max,2, min,max,Bar,min,max

Returns the ranges of the installed
sensors for the active channel.

Localgravity

Value in ft/s²

Sets the local gravity in feet/sec².

Localgravity?

<sp>n.nnnnne+nn<cr><lf>

Returns the local gravity in feet/sec².

LowerLimit

Value inside primary xducer Range
on turndown #1 in current units.

Sets the lower control limit for the
instrument.

LowerLimit?

<sp>xxxxxxx<cr><lf>

Returns the lower control limit for the
instrument in current units.

Measure

None

Instrument placed in Measure Mode.

Measure?

<sp>(Yes or No)<cr><lf>

Returns YES if instrument is in
measure, No if otherwise.

Netmask

nnn.nnn.nnn.nnn

Sets the Ethernet network mask.

Netmask?

<sp>nnn.nnn.nnn.nnn<cr><lf>

Returns the Ethernet network mask.

Outform

1 to 8 – see table below

Sets the output format.

Outform?

<sp>X<cr><lf>

Returns the output format – see table
below

Peakmax?

<sp>n.nnnnne+nn<cr><lf>

Returns the maximum pressure since
peakreset was sent.

Peakmin?

<sp>n.nnnnne+nn<cr><lf>

Returns the minimum pressure since
peakreset was sent.

Peakreset

None

Resets the peak values.

Port

nnnnnn

Sets the Ethernet port of the instru-
ment

Port?

<sp>nnnnn<cr><lf>

Returns the Ethernet port of the instru-
ment

Ptype

Absolute or Gauge

Sets the instrument pressure type
– gauge only works if the optional
barometric sensor is installed

Ptype?

<sp>CCCCC<cr><lf>

Returns “Absolute” or “Gauge” for the
pressure type

RangeMax?

<sp>XXXXXXX<cr><lf>

Returns the maximum range of the
active transducer and turndown in the
current units.

RangeMin?

<sp>XXXXXXX<cr><lf>

Returns the minimum range of the
active transducer and turndown in the
current units.

Rate?

<sp>XXXXXXX<cr><lf>

Returns the rate reading of the instru-
ment in current units/second.