Pneumatic High-Speed Pressure Controller CPC3000

WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1

55

11498171.01 08/2009 GB

Command

Response/Function

CALCulate

:LIMit

:LOWer <n>

Set the minimum control limit

:LOWer?

Set the minimum control limit

:UPPer <n>

Set the maximum control limit

:UPPer?

Set the maximum control limit

:SYSTem

:DETECT SLOW | FAST | CANCEL

Control autotune (NOT USED NOW)

:DETECT?

Returns state of control autotune

Output formats
1. <sp> pressure value <cr><lf>
2. <sp> pressure, units number, mode <cr><lf>
3. <sp> pressure, pressure rate <cr><lf>
4. <sp> pressure, minimum peak, maximum peak<cr><lf>
5. <sp> pressure, active sensor (P or S) active turndown (1-4)<cr><lf>
6. <sp> pressure, control point, “stable” or “slewing”<cr><lf>
7. <sp> pressure, “no barometer” or baro reading<cr><lf>