8. Commissioning, operation

With safety pressure gauges (see dial symbol k) ensure that the free

space behind the blow-out back is at least 15 mm.

8.2 Special requirements for the installation point

The instruments must be mounted in a location free from vibration in order to

avoid contact chatter and the resulting increased wear.

If the line to the measuring instrument is not adequately stable, a measuring

instrument holder should be used for fastening (and possibly via a flexible

capillary).

If vibrations cannot be avoided by means of suitable instal-

lation, instruments with liquid filling should be used. The instruments

should be protected against coarse dirt and wide fluctuations in ambient

temperature.

8.3 Electrical connection
The electrical connection must only be made by qualified skilled

personnel. Connection details and switching functions are given on

the product label. Connection terminals and earth terminal are appro-

priately marked. The mains connection lines to be provided must be

dimensioned for maximum instrument power consumption and comply

with IEC 227 or IEC 245. Power ratings (see "Specifications") and

protective circuits with inductive and capacitive loads see chapter

"9. Contact protection measures".