Notice:
In order to carry out a product control using “non-destructive average tare”
mode, in accordance with the regulation on standard deviation “S”, the mass
of packages is determined from at least 10 measurements, and the standard
deviation can not be greater than 0.25 of the maximum permissible negative
error T for nominal mass of a package.
If an operator carries out the last measurement of the packaging mass, the
software displays a summary of the process. Automatically, the system
generates a report is saved in the scale database:
Pressing
key causes moving to the control process without saving the
determined average mass of a packaging in product record in the database.
Pressing
key causes moving to the control process with simultaneous
saving the determined average mass of a packaging in product record in the
database.
During the control process in progress the software carries out real-time
updating of measurement results and displays them in the corresponding
fields for informing an operator on control result: