RADWAG WLY Scales User Manual

Page 137

Advertising
background image

137

Notice:
In order to carry out a product control using “non-destructive average tare”
mode, in accordance with the regulation on standard deviation “S”, the mass
of packages is determined from at least 10 measurements, and the standard
deviation can not be greater than 0.25 of the maximum permissible negative
error T for nominal mass of a package.

If an operator carries out the last measurement of the packaging mass, the

software displays a summary of the process. Automatically, the system

generates a report is saved in the scale database:

Pressing

key causes moving to the control process without saving the

determined average mass of a packaging in product record in the database.

Pressing

key causes moving to the control process with simultaneous

saving the determined average mass of a packaging in product record in the

database.

During the control process in progress the software carries out real-time

updating of measurement results and displays them in the corresponding

fields for informing an operator on control result: