RADWAG WPW 60/H5/K Multifunctional Scales User Manual

Page 123

Advertising
background image

When inscribing a reference quantity, remember that the unit mass Cannot
be lower than 0.1 of reading division and the total reference mass Carnot
be lower than 1 reading division. If these conditions are not fulfilled an
error message will be displayed.

• After confirming the inscribed value by

, the program

will display:

If pieces are weighted in a container, it should be put on the pan
and tarred and then put the declared value of the reference quantity.
When stable (

) confirm the measurement.

Software will automatically calculate the unit mass and return to counting
pieces with indication of number of pieces:

Caution:

When the reference mass is confirmed by

and the pan is empty,

the display will show <Err6> and the program return to the previous step.

If <P7.5.8 AUTO COR.> parameter is set to YES program will activate
the function of Automatic Accuracy Correction.

The reference quantity can be increased by adding subsequent pieces.
Number of pieces and unit mass (SMP) will be automatically corrected
by software and show in the bottom line of the display:

- 123 -