28.4.6. Send the result immediately in basic unit
Syntax: SI CR LF
Possible answers:
SI_I CR LF
- command comprehended but cannot be executed at the
moment
SI_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in basic unit is returned
Frame format:
1 2
3
4
5
6 7-15
16 17 18 19 20 21
S I space stability space sign mass space
unit
CR LF
Example:
S I CR LF – computer command
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass value
in basic unit is returned immediately.
28.4.7. Get immediate results from both platforms in basic units
Syntax: SIA CR LF
Possible answers:
SIA_I CR LF
- command comprehended but cannot be executed at the moment
MASS FRAME „P1” CR LF
MASS FRAME „P2” CR LF
- mass values are immediately returned from both
platforms in basic units
Frame format with mass from subsequent platforms as indicator reply:
1 2
3
4
5
6 7-15 16 17 18 19 20 21
P n space stability space
sign
mass
space
unit
CR
LF
- WPW 60/H4/K Multifunctional Scales WPW 60/H3/K Multifunctional Scales WPW 60/C2/K Multifunctional Scales WPW 6/H2/K Multifunctional Scales WPW 6/F1/K Multifunctional Scales WPW 300/H6/K Multifunctional Scales WPW 300/H5/K Multifunctional Scales WPW 300/C3/K Multifunctional Scales WPW 300/C2/K Multifunctional Scales WPW 30/H3/K Multifunctional Scales WPW 30/F1/K Multifunctional Scales WPW 30/C2/K Multifunctional Scales WPW 3/H1/K Multifunctional Scales WPW 150/H6/K Multifunctional Scales WPW 150/H5/K Multifunctional Scales WPW 150/H4/K Multifunctional Scales WPW 150/H3/K Multifunctional Scales WPW 150/C3/K Multifunctional Scales WPW 150/C2/K Multifunctional Scales WPW 15/H3/K Multifunctional Scales WPW 15/H2/K Multifunctional Scales WPW 15/F1/K Multifunctional Scales