28.4.9. Send the result immediately in current unit
Syntax: SUI CR LF
Possible answers:
SUI_I CR LF
- command comprehended but cannot be executed
SUI_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in current unit is returned immediately
Frame format:
1 2 3
4
5
6 7-15 16 17 18 19 20 21
S U I stability space sign mass space
unit
CR LF
Example:
S U I CR LF – computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed and mass
returned
28.4.10. Switch on continuous transmission in basic unit
Syntax: C1 CR LF
Possible answers:
C1_I CR LF
- command comprehended but cannot be executed
C1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1 2
3
4
5
6 7-15
16 17 18 19 20 21
S I space stability space sign mass space
unit
CR LF
- WPW 60/H4/K Multifunctional Scales WPW 60/H3/K Multifunctional Scales WPW 60/C2/K Multifunctional Scales WPW 6/H2/K Multifunctional Scales WPW 6/F1/K Multifunctional Scales WPW 300/H6/K Multifunctional Scales WPW 300/H5/K Multifunctional Scales WPW 300/C3/K Multifunctional Scales WPW 300/C2/K Multifunctional Scales WPW 30/H3/K Multifunctional Scales WPW 30/F1/K Multifunctional Scales WPW 30/C2/K Multifunctional Scales WPW 3/H1/K Multifunctional Scales WPW 150/H6/K Multifunctional Scales WPW 150/H5/K Multifunctional Scales WPW 150/H4/K Multifunctional Scales WPW 150/H3/K Multifunctional Scales WPW 150/C3/K Multifunctional Scales WPW 150/C2/K Multifunctional Scales WPW 15/H3/K Multifunctional Scales WPW 15/H2/K Multifunctional Scales WPW 15/F1/K Multifunctional Scales