IKA KS 4000 ic control User Manual

Page 21

Advertising
background image

­­21

KS 4000 i 1208

Tapping the Start/Stop keys or On/Off key starts the particular fun-
ctions

The desired default values can be changed using the s or t keys.

Use the shift key 1 to switch from hour/minute mode to minu-
te/second mode.

If the elapsed time exceeds the value of 100 hours, the display

switches from hour/minute mode

to hour mode

Only whole hours are displayed in hour mode.

The following sections contain detailed information about the indi-
vidual functions.

Press the 3 key to start or stop the shake function. The speed
can be adjusted during operation. The displayed value flashes until
the pre-set speed has been reached.

When the shaking function is started using button 3 , the timer
automatically starts measuring the time until the next switch-off.
Note:
The shaking function can only be started when the cover is closed.
When the cover is opened, the shaker automatically switches off
the shaking and heating functions.

Function Shaking

The desired shaking duration is set on the operator panel of the
timer using the Time up/down keys. The shift key can be used to
switch from hour mode to minute mode.

The Time Start/Stop key activates the timer and shaking functions.

When the time is up, the shaking function stops and an acoustic
signal is given off. If the cover is opened before the time is up, the
timer goes to "pause" and the display flashes. If the cover remains
open for more than 15 minutes, the timer switches off and an
error code appears on the display. The shaker no longer starts
when the cover is closed.

If the timer is not activated, the Time display automatically shows
the operating time following the start of the shaking function.

Once the appliance starts, the display indicates the actual value for
the set temperature.

Press the 4 key to start or stop the heating function. The tempe-
rature is entered using the Temp up/down keys. In operation, the
temperature is displayed in 0.1 °C steps.

The target value appears approx. every 5 seconds and remains on
the display for 2 seconds.

Note:
The heating function can only be started when the cover is closed.
When the cover is opened, the shaker automatically switches off
the shaking and heating functions.

Timer