Reset, Appliance variant ks 4000 ic control – IKA KS 4000 ic control User Manual

Page 23

Advertising
background image

23

KS 4000 i 1208

Calibration without inserted sensor:
- Fill Erlenmeyer flask (250ml) with water to the 100ml mark
- Place the Erlenmeyer flask in the centre
- Immerse the sensor of the external measuring device in the

water

- Set the target temperature
- Close the cover and start the Temp function
- Wait until the temperature in the incubation cover has stabilised.
- While holding down the shift key 1 , briefly press the Temp

On/Off key 4 , the

CAL

display appears; keep holding down

the shift key 1 .

- Set the temperature value read off the external measuring devi-

ce using the Temp up s down t keys

- Calibration is complete when you release the shift key

Calibration with inserted sensor:
Calibration with inserted sensor takes place in the same way as
described above.

To restore the factory settings, hold down the Mot t key and the
Temp t key and press the main switch I/O.

Factory setting:

Operating mode A
Upper speed limit 500 rpm
Safety temperature limit 90 °C
Temperature calibration

Reset

There is a cooler built into this appliance variant.

By using an external cooling unit, the working temperature can be
lowered in relation to the room temperature (depending on the
supply temperature).
The cooler is connected to an external cooling unit (e.g. IKA KV
600) via the plug connection at the back of the appliance. The inlet
and outlet connections are labelled accordingly at the back of the
appliance. To connect the cooling pipes there are two hose
connection pieces included. They can be connected to a hose with
a 10mm interior diameter. The connectors are unlocked by coaxi-
ally positioning the unlocking lever and pressing slightly in the
direction of the arrow. By positioning and light pressure in the
direction of the plug, the connectors are locked/connected to the
inlet/outlet plug connections on the housing.

Appliance variant KS 4000 ic control

Drain hose

Plug connection

cooler outlet

Plug connection

cooler inlet

Unlocking lever