IKA KS 4000 ic control User Manual

Page 25

Advertising
background image

25

KS 4000 i 1011

Error code Description

Cause

Effect

Solution

Er 3

Temperature inside unit too high

• Permitted ambient temperature has

Heating off

• Switch off the unit. Allow it to cool down

been exceeded

and then switch on again.

• Ventilation slots or fan housing

• Clean ventilation slots or fan

blocked

housing

• Observe maximum permissible

ambient temperature

Er 4

Differanceberween setpoint and

• Motor seized or overloaded

Heating off

• Reduce the torque load

actual speeds is too large

(load)

• Reduce the setpoint speed

Er 8

The calibration value of the tempe- • Fault in calibration procedure

Heating off

• Repeat the calibration procedure

rature sensor is outside the limit

• Value was incorrectly stored to memory

value

• EPROM switching error

Er 14

External temperature sensor,

• Short-circuit at temperature sensor

Heating off

• Check the plug

short-circuit

plug

• Substitude the temperature sensor

• Short-circuit in connecting cable

or on temperature sensor

Er 16

External temperature sensor has

• SAFE Temp is set at a temperature

Heating off

• Allow the unit to cool

exceeded the SAFE Temp

lower than the actual temperature on

• Set the SAFE Temp at a higher temperature

the external temperature sensor

Er 17

Temperature sensor - Incubation-

• SAFE Temp is set at a temperature

Heating off

• Allow the unit to cool

room has exceeded the

lower than the actual temperature on

• Set the SAFE Temp at a higher temperature

SAFE Temp

the external temperature sensor

Er 26

Difference between the internal

• Ventilation slots in incubation room

Heating off

• Switch off the unit. Allow it to cool down

contro and safety temperature

blocked

and then switch on again.

sensors too large

• Radial-flow fan does not rotaded

• Check fan or ventilation slots and

• Fault in the control or safety

clean if necessary

temperature sensors

Er 60

Power outage

• Power outage during operation

Interruption of heating

• Delete the display by pressing the

or shaking function

shift key

Advertising