IKA KS 4000 ic control User Manual

Page 22

Advertising
background image

­­22

Locate the external
temperature

sensor

and the connecting
cable so that the sha-
king motion is unimpeded and no vessels are damaged or
tipped over.

KS 4000 i 1208

You can choose between operating modes

and

Appliance does not start up again following power
outage Safety limit value for speed and temperature
adjustable

Appliance starts up again following power outage
Safety limit value for speed and temperature adjustable

Appliance starts up again following power outage
Safety limit values cannot be changed

Switch on appliance and simultaneously hold down the Time
Start/Stop key 2 and the shift key 1 .

The display indicates either or depending on the opera-
ting mode set (factory default setting A).

Switching the appliance off and on switches to the next operating
mode

Following the selection of the operating mode, the corresponding
letter appears on the display for 5 seconds, the appliance is ready
for operation.

Operating modes

Setting the operating mode

If external temperature sensor PT1000.60 is connected to the
internal slide-on receptacle, the temperature can be measured at
any place in the medium.

(Connector for PT1000.60)

The temperature control of the
heating works automatically with
this display or measuring value.
This is indicated by the lighting up
of the LED next to the temperatu-
re sensor symbol

External temperature sensor

The appliance is calibrated at the factory.

This function allows the temperature to be calibrated to a desired
value.
This may be necessary, for example, if special sample containers,
their arrangement or external influences (e.g. sunshine) cause the
measuring value for the temperature to be incorrect.

Calibration - temperature

Advertising
This manual is related to the following products: