IKA KS 4000 ic control User Manual

Page 29

Advertising
background image

29

KS 4000 i 1011

approved by IKA to clean IKA devices.
To remove use:
Dyes

isopropyl alcohol

Construction materials water containing tenside / isopropyl alcohol
Cosmetics

water containing tenside / isopropyl alcohol

Foodstuffs

water containing tenside

Fuels

water containing tenside

For materials which are not listed, please request information
from IKA. Wear the proper protective gloves during cleaning of
the devices.
Electrical devices may not be placed in the cleansing agent for
the purpose of cleaning.
Before using another than the recommended method for clea-
ning or decontamination, the user must ascertain with the
manufacturer that this method does not destroy the instru-
ment.

Collecting tray with drain hose (for picture see below)
(applies to KS4000 i control and KS4000 ic control)
In the event of glass breakage, leaking liquid is collected by the
collecting tray below the shaking table and guided to the drain
hose via a drain at the rear.

1. Remove drain hose

from bracket

2. Place drain hose

in laboratory drain

Cleaning of the collec-
ting tray:

- four countersunk

head screw at the
top side of the
shaking table
loosen

- shaking table

upward remove

Cleaning the Plexiglas cover:
- Do not dry wipe.
- Do not use abrasive materials.
- Do not use solvents.

Clean dusty surfaces with warm water, detergent and a soft
cloth.
We recommend the anti-static plastic cleaner “AKU” by Burnus
GmbH, Darmstadt.

For disinfecting, only use products prescribed by the manufac-
turer specially for use on Plexiglas.