12 mode dilute and mix – mixing liquid, Mode dilute and mix – mixing liquid – Eppendorf epMotion 96 User Manual
Page 46

Operation
epMotion
®
96
English (EN)
46
6.12
Mode
Dilute and Mix
– Mixing liquid
The application is suitable for diluting samples and reagents using an appropriate dilution solution.
Aspirate the concentrated liquid, aspirate diluent and mix while dispensing.
Abb. 6-5: Mode Dilute and Mix
Fig. 6-5:
Mode
Dilute and Mix
Parameter
Value
Range of values
Speed aspirate
Setting the speed of the liquid aspiration.
1 – 9
Speed dispense
Setting the speed of the liquid dispensing.
1 – 9
Diluent Vol.
Setting the aspiration volume for the diluent in
μL.
0.5 – 300 (increment of 0.1)
Air Gap Vol.
Setting the volume of the air gap in μL.
0.5 – 300 (increment of 0.1)
Concentrate Vol.
Setting the aspiration volume of the concentrate
in μL.
0.5 – 300 (increment of 0.1)
Mix vol.
Setting the mixing volume in μL.
0.5 – 300 (increment of 0.1)
Mix Cycles
Setting the number of mixing cycles.
1 – 99