6.15.5
Changing the liquid type
Liquid properties are defined for individual pipette tip sizes. Calibration values may be entered for a
maximum of 5 different volume ranges. The higher the calibration value, the greater the piston stroke per
microliter of the set pipetting volume.
The calibration values of the liquids must be determined empirically. This can be done using a
gravimetric comparison measurement, for example.
It is not possible to change the
Standard
liquid type.
1. Select a liquid type and confirm with
Select
.
The system opens the
Liquid Type Editor
.