2 test instructions, upper fluorescence, Test instructions, upper fluorescence – Eppendorf AF2200 Plate Reader User Manual
Page 51

Maintenance
Eppendorf
®
PlateReader AF2200
English (EN)
51
8.6.2
Test instructions, upper fluorescence
The following tests can be conducted on the device to check the device specifications for this measuring
procedure:
• Sensitivity
• Precision within a plate
• Precision from one measurement to the next
Sensitivity
Conduct the measurement with the following settings to determine the detection limit for fluorescein.
Pipette 200 μL of 1 nM fluorescein or the blank solution (0.01 M NaOH) into the corresponding wells
according to the plate layout below.
Parameter
Setting
Measuring mode
Upper fluorescence
Wavelength, excitation
485 (20 nm)
Wavelength, emission
535 (25 nm)
Number of flashes
25
Integration time
40 (μs)
Setting time
0 (μs)
Signal gain
Optimal
Plate type
GRE96fb
X
1 nM fluorescein
blank
0.01 M NaOH