Eppendorf AF2200 Plate Reader User Manual
Page 54

Maintenance
Eppendorf
®
PlateReader AF2200
English (EN)
54
Precision from one measurement to the next
Carry out the following measurement to determine the precision and reproducibility.
Pipette 200 μL of 1 nM fluorescein or the blank solution (0.01 M NaOH) into the corresponding wells
according to the plate layout below.
Material/reagents
• 1 nM fluorescein (in 0.01 M NaOH) (fluorescein sodium salt)
• 0.01 M NaOH (=blank solution) (NaOH pellets)
• 96-well plate
• 200 μL pipette with tips
Parameter
Setting
Measuring mode
Upper fluorescence
Wavelength, excitation
485 (20 nm)
Wavelength, emission
535 (25 nm)
Number of flashes
25
Integration time
40 (μs)
Setting time
0 (μs)
Signal gain
Optimal
Plate type
GRE96fb
Part of plate
A1
Kinetic
20 cycles
Time interval
Minimum
X
1 nM fluorescein
blank
0.01 M NaOH