Eppendorf AF2200 Plate Reader User Manual
Page 57
Advertising

Maintenance
Eppendorf
®
PlateReader AF2200
English (EN)
57
Precision across the plate
Carry out the following measurement to determine the precision across the plate.
Pipette 200 μL of 25 nM fluorescein or the blank solution (0.01 M NaOH) into the corresponding wells
according to the plate layout below.
Material/reagents
• 25 nM fluorescein (in 0.01 M NaOH) (fluorescein sodium salt)
• 0.01 M NaOH (=blank solution) (NaOH pellets)
• 96-well plate
• 200 μL pipette with tips
Parameter
Setting
Measuring mode
Lower fluorescence
Wavelength, excitation
485 (20 nm)
Wavelength, emission
535 (25 nm)
Number of flashes
25
Integration time
40 (μs)
Setting time
0 (μs)
Signal gain
Optimal
Plate type
GRE96fb
X
25 nM fluorescein
blank
0.01 M NaOH