5.3.2.3 Measuring the blank and standards
5.3.2.4 Measuring samples
For evaluations without standards (e.g. DNA measurements), this method step is omitted.
1. Start by measuring a blank (
blank key).
2. Then measure all standards one by one
(
standard key).
The display always marks the standard that is to be
measured next. Use the [Graph] resp. [Table]
softkey to change the result view.
Press [Next] to accept the evaluation calculated
from the standard results.
The sample key is used for measuring your
samples consecutively.
Blank results will remain saved for the duration of
one series of measurement. However, a new blank
measurement always is possible. (The adjacent
figure shows a measuring procedure with
evaluation via the standard curve and, in addition
to the sample result, the graph of the standard
evaluation.)