Eppendorf BioSpectrometer kinetic User Manual
Page 99

99
Evaluation procedure
Eppendorf BioSpectrometer
®
kinetic
English (EN)
When the kinetic method is evaluated with standards the measurement of a reagent blank is not available.
Here, the reagent blank can alternatively be defined as the first standard ("zero" concentration).
If the time frame for the kinetic evaluation with linear regression is reduced in the
process
results method step, the reduced measuring window is automatically used for calculating the
reagent blank which was incorporated in the calculation of this sample result. That means that
the reagent blank is recalculated (only) for the calculation of this sample result. For the
calculation of the other sample results the reagent blank result that was calculated on the
basis of the originally used time window is used.